Search In this Thesis
   Search In this Thesis  
العنوان
A manufacturer-independent technique for ultrasound elastography using conventional grayscale images /
الناشر
Mohamed Tarek Mohamed Ali ,
المؤلف
Mohamed Tarek Mohamed Ali
هيئة الاعداد
باحث / Mohamed Tarek Mohamed Ali
مشرف / Ahmed Sohier Fahmy
مشرف / Ahmed Mohamed Ehab Mahmoud
مشرف / Ahmed Sohier Fahmy
تاريخ النشر
2017
عدد الصفحات
71 P. :
اللغة
الإنجليزية
الدرجة
ماجستير
التخصص
الهندسة الطبية الحيوية
تاريخ الإجازة
3/3/2018
مكان الإجازة
جامعة القاهرة - كلية الهندسة - Systems and Biomedical Engineering
الفهرس
Only 14 pages are availabe for public view

from 90

from 90

Abstract

Ultrasound strain elastography has been widely used to assess tissue mechanical properties. It provides a complementary tool to diagnose tumors and cardiovascular diseases. However, ultrasound elastography is only available on high-end ultrasound machines, and/or requires an access to the radio frequency (RF) raw ultrasound signals that are not available on most ultrasound machines. This wok developed and assessed a novel manufacturer-independent elastography ultrasound technique that uses conventional grayscale ultrasound images to calculate strain elastography. The technique can be integrated with commercially available ultrasound machines to reconstruct strain using freehand quasi-static compression